证:工商信息已核宝br /> 访问量:86141
日本光明理化?02SA丙酮气体检测管可以检?种量程,0.1%~2%?%~5%
用途:
用于检测空气的丙酮浓度,并可利用换算刻度检测四氢呋喃的浓度、/span>
102SA丙酮检测管说明
用途:
用于检测空气的丙酮浓度,并可利用换算刻度检测四氢呋喃的浓度、/span>
规格9/span>
样本采集野/span> 咋/span> 检测范図/span> |
检测气体名?/span> |
100ml |
50ml(注1(/span> |
丙酮 |
0.1~2.0%(试管刻度) |
1.0~5.0%(换算刻度I) |
|
四氢呋喃 |
0.2~3.0%(换算刻度II) |
0.1~2.0%(试管刻度III) |
|
检测时闳/span> |
1.5分钟/100ml |
||
检测限?/span> |
200ppm(丙酮) 20ppm(四氢呋喂strong>(/strong> |
||
检测剂变色 |
桔色→深棕色 |
||
使用温度范围 |
0~40ℂ/span>(需温度补正(/span> |
||
温度影响 |
旟/span> |
||
反应原理 |
氧化铬被还原 |
干扰气体
气体同/span> |
?%) |
哌/span> |
酒精籺/span> |
指示偏高 |
|
籺/span> |
指示偏高 |
|
籺/span> |
指示偏高 |
|
芳香族碳化氢 |
指示偏高 |
|
卤化碳化氡/span> |
0.5 |
指示偏高 |
检测顺庎/span>用切割槽将检测管的两端折掉、/span>
①/span>将检测管上的箭头'/span>G↑/span>)朝向采集器并安装在上面、/span>
将拉手推入到*里面,使拉手与管身上红色的标记对上、/span>
③/span>将拉手迅速拔出到底,拉手被固定、/span>
⑤抽?00ml气体时手泵的红色指示器弹出就说明气体已全部进入检测管?/span>
取下检测管+/span>根据变色层的长度读取刻度、/span>
?-/span>当需采集50ml时,②/span>~④/span>的操作共要进衋/span>4次。温度补正值要乘以0.2、/span>
?/span>2.因本检测管所含水份较多,为防止采集器生锈、泄漏,使用后请在采集上不安装任何东西的状态下,将拉手来回拈/span>5次、/span>
换算刻度
换算刻度I(采雅/span>50ml丙酮(/span>
换算刻度II(采雅/span>100ml四氢呋喃(/span>
换算刻度III(采雅/span>50ml四氢呋喃(/span>
浓度单位
必要时根据下面的公式换算浓度单位9/span>
测量浓度'/span>mg/m3(/span>=测量浓度(%?#215;分子野/span>/22.4273/(273+t)10000
t9/span>:测定时的室温
检测气体名?/span> |
丘/span>?/span> |
四氢呋喃 |
刅/span>孏/span>野/span> |
58.08 |
72.11 |
使用环境
温度9/span>20ℂ/span>以外的环境时+/span>利用下面的温度补正表进行补正、/span>
丙酮的温度补偿表(20℃标?
读取倻/span> (%(/span> |
真实的浓度(%) |
||||
0ℂ/span> |
10ℂ/span> |
20ℂ/span> |
30ℂ/span> |
40ℂ/span> |
|
2.0 |
,/span> |
2.38 |
2.00 |
1.78 |
1.60 |
1.5 |
2.20 |
1.76 |
1.50 |
1.30 |
1.16 |
1.0 |
1.44 |
1.18 |
1.00 |
0.86 |
0.76 |
0.5 |
0.72 |
0.60 |
0.50 |
0.42 |
0.36 |
0.2 |
0.30 |
0.25 |
0.20 |
0.16 |
0.14 |
0.1 |
0.16 |
0.12 |
0.10 |
0.08 |
0.08 |
四氢呋喃温度补偿?/span>(20℃标?
读取倻/span> (%(/span> |
真实的浓度(%) |
||||
0ℂ/span> |
10ℂ/span> |
20ℂ/span> |
30ℂ/span> |
40ℂ/span> |
|
5.0 |
6.0 |
5.3 |
5.0 |
4.8 |
4.5 |
4.0 |
4.8 |
4.3 |
4.0 |
3.8 |
3.6 |
3.0 |
3.6 |
3.2 |
3.0 |
2.9 |
2.7 |
2.5 |
3.0 |
2.7 |
2.5 |
2.4 |
2.3 |
2.0 |
2.4 |
2.1 |
2.0 |
1.9 |
1.8 |
1.5 |
1.8 |
1.6 |
1.5 |
1.4 |
1.4 |
1.0 |
1.1 |
1.1 |
1.0 |
1.0 |
0.9 |
0.5 |
0.6 |
0.5 |
0.5 |
0.5 |
0.5 |
湿度:无影响、/span>
气压:使用下面的公式进行气压补正、/span>
温度补正后的值(ppm(/span>1013'/span>hPa(/span>检测点的气压(hPa(/span>